Remote Commands
R&S
®
FSW
1301User Manual 1173.9411.02 ─ 43
DIAGnostic:INFO:CCOunt? <Relay>
This command queries how many switching cycles the individual relays have per-
formed since they were installed.
Query parameters:
<Relay> ACDC
Mechanical Attenuation Coupling
ATT5
Mechanical Attenuation 05 DB
ATT10
Mechanical Attenuation 10 DB
ATT20
Mechanical Attenuation 20 DB
ATT40
Mechanical Attenuation 40 DB
CAL
Mechanical Calibration Source
EATT
Electrical Attenuation Bypass
PREamp
Preamplifier Bypass
Return values:
<Cycles> Number of switching cycles.
Example:
DIAG:INFO:CCO? CAL
Usage: Query only
Manual operation: See "Relays Cycle Counter" on page 719
DEVice:INFO:HWBand?
Queries the frequency bands used for measurement by the R&S FSW hardware. The
start frequency of each band is provided.
The bands are instrument-specific and depend on the currently defined RBW, VBW
and YIG preselector state. The precise frequency bands are required to define correc-
tion data for the correct bands, in particular for frequency-drifting DUTs.
This query is only available in zero span mode.
Return values:
<StartFreq>
Configuring the R&S FSW