Remote Commands
R&S
®
FSW
1034User Manual 1173.9411.02 ─ 43
14.7.2.2 Configuring the Sweep
Useful commands for configuring sweeps described elsewhere:
[SENSe:]AVERage<n>:COUNt on page 1129
[SENSe:]AVERage<n>[:STATe<t>] on page 1129
[SENSe:]AVERage<n>:TYPE on page 1129
Remote commands exclusive to configuring sweeps:
[SENSe:]SWEep:COUNt...............................................................................................1034
[SENSe:]SWEep:DURation?......................................................................................... 1035
[SENSe:]SWEep:FFTSubspan?.....................................................................................1035
[SENSe:]SWEep:OPTimize...........................................................................................1035
[SENSe:]SWEep[:WINDow<n>]:POINts..........................................................................1036
[SENSe<n>:]SWEep:TIME............................................................................................1037
[SENSe:]SWEep:TIME:AUTO....................................................................................... 1037
[SENSe:]SWEep:TYPE.................................................................................................1038
[SENSe:]SWEep:TYPE:USED.......................................................................................1038
[SENSe:]BWIDth[:RESolution]:FFT................................................................................1038
[SENSe:]BANDwidth[:RESolution]:FFT...........................................................................1038
[SENSe:]SWEep:COUNt <SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In continuous sweep mode, the application calculates the moving average over the
average count.
In single sweep mode, the application stops the measurement and calculates the aver-
age after the average count has been reached.
Parameters:
<SweepCount> When you set a sweep count of 0 or 1, the R&S FSW performs
one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range: 0 to 200000
*RST: 0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation: See "Sweep/Average Count" on page 470
Setting Basic Measurement Parameters

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals