General Instrument Setup
R&S
®
FSW
682User Manual 1173.9411.02 ─ 43
Unit
The unit in which the y-values of the data points of the transducer line are defined.
As soon as a transducer is activated, the unit of the transducer is automatically used
for all the level settings and outputs. The unit cannot be changed in the amplitude set-
tings unless dB is used.
Remote command:
[SENSe:]CORRection:TRANsducer:UNIT on page 1270
X-Axis Scaling
Describes the scaling of the horizontal axis on which the data points of the transducer
line are defined. Scaling can be linear or logarithmic.
Remote command:
[SENSe:]CORRection:TRANsducer:SCALing on page 1269
Data Points
Each transducer line is defined by a minimum of 2 and a maximum of 50 data points.
Each data point is defined by its position (x-axis) and value (y-value).
The data points must comply with the following rules to ensure correct operation:
The frequencies for the data points must always be defined in ascending order.
Otherwise the entry will not be accepted and the an error message is displayed.
The frequencies of the data points may exceed the valid frequency range of the
R&S FSW since only the set frequency range is taken into account for measure-
ments. The minimum frequency of a data point is 0 Hz, the maximum frequency
200 GHz.
The value range for the transducer factor is ±200 dB.
Gain has to be entered as a negative value, and attenuation as a positive value.
Remote command:
[SENSe:]CORRection:TRANsducer:DATA on page 1269
Insert Value
Inserts a data point in the transducer line above the selected one in the "Edit Trans-
ducer" dialog box.
Delete Value
Deletes the selected data point in the "Edit Transducer" dialog box.
Shift x
Shifts the x-value of each data point horizontally by the defined shift width.
Shift y
Shifts the y-value of each data point vertically by the defined shift width.
Save
Saves the currently edited transducer line under the name defined in the "Name" field.
Remote command:
MMEMory:SELect[:ITEM]:TRANsducer:ALL on page 1237
MMEMory:STORe<1|2>:STATe on page 1240
Transducers