Remote Commands
R&S
®
FSW
1313User Manual 1173.9411.02 ─ 43
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 720
INSTrument:COUPle:BANDwidth <State>
INSTrument:COUPle:BWIDth <State>
This command turns synchronization of the resolution bandwidth (and filter type)
between measurement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns on synchronization.
*RST: NONE
Example:
INST:COUP:BWID ALL
Synchronizes the resolution bandwidth.
INSTrument:COUPle:CENTer <State>
This command turns synchronization of the frequency between measurement channels
on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: ALL
Example:
INST:COUP:CENT ALL
Synchronizes the center frequency.
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 720
INSTrument:COUPle:DEMod <State>
This command turns synchronization of the audio demodulator configuration between
measurement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
Configuring the R&S FSW

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals