Remote Commands
R&S
®
FSW
1314User Manual 1173.9411.02 ─ 43
*RST: NONE
Example:
INST:COUP:DEM ALL
Synchronizes the audio demodulator configuration.
INSTrument:COUPle:GAIN <State>
This command turns synchronization of the preamplifier configuration between mea-
surement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:GAIN ALL
Synchronizes the preamplifier configuration.
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 720
INSTrument:COUPle:IMPedance <State>
This command turns synchronization of the impedance configuration between mea-
surement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:IMP ALL
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 720
INSTrument:COUPle:LIMit <State>
This command turns synchronization of limit results between measurement channels
on and off.
Parameters:
<State> ALL | NONE
Configuring the R&S FSW