Remote Commands
R&S
®
FSW
1315User Manual 1173.9411.02 ─ 43
ALL | 1
Turns on synchronization.
Limit lines have to be compatible to the x-axis and y-axis config-
uration for successful synchronization.
NONE | 0
Turns off synchronization.
*RST: ALL
Example:
INST:COUP:LIM ALL
Synchronizes the limit values.
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 720
INSTrument:COUPle:LLINes
<State>
This command turns synchronization of the limit lines between measurement channels
on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
INSTrument:COUPle:MARKer <State>
This command turns synchronization of the marker frequency between measurement
channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:MARK ALL
Synchronizes the receiver frequency and the marker frequency.
INSTrument:COUPle:PRESel <State>
This command turns synchronization of the preselector state between measurement
channels on and off.
Parameters:
<State> ALL | NONE
Configuring the R&S FSW