Remote Commands
R&S
®
FSW
1164User Manual 1173.9411.02 ─ 43
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See "Search Next Peak" on page 530
CALCulate<n>:MARKer<m>:MINimum:AUTO
<State>
This command turns an automatic marker peak search for a trace minimum on and off.
The R&S FSW performs the peak search after each sweep.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:MIN:AUTO ON
Activates the automatic minimum value search function for
marker 1 at the end of each particular sweep.
Manual operation: See "Auto Max Peak Search / Auto Min Peak Search"
on page 525
CALCulate<n>:MARKer<m>:MINimum:LEFT
This command moves a marker to the next minimum value.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See "Search Next Minimum" on page 531
Analyzing Measurements (Basics)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals