EasyManua.ls Logo

R&S FSW Series - Page 1122

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
1122User Manual 1173.9411.02 ─ 43
1 = origin of coordinate system (x1 = 0, y1 = 0)
2 = end point of system (x2 = 100, y2= 100)
3 = zoom area (e.g. x1 = 60, y1 = 30, x2 = 80, y2 = 75)
Suffix:
<n>
.
Window
<w> subwindow
Parameters:
<x1> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
<y1> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
<x2> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
<y2> Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
Manual operation: See "Single Zoom" on page 510
DISPlay[:WINDow<n>][:SUBWindow<w>]:ZOOM[:STATe] <State>
This command turns the zoom on and off.
Suffix:
<n>
.
Window
<w> subwindow
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
Analyzing Measurements (Basics)