Measurements and Results
R&S
®
FSW
201User Manual 1173.9411.02 ─ 43
Figure 7-35: Adjacent channel power measurement on a CDMA2000 signal
7.2.7.2 Measurement Example 2 – Measuring Adjacent Channel Power of a W-CDMA
Uplink Signal
Test setup:
Signal generator settings (e.g. R&S SMW):
Frequency: 1950 MHz
Level: 4 dBm
Modulation: 3GPP W-CDMA Reverse Link
Procedure:
1. Preset the R&S FSW.
2. Set the center frequency to 1950
MHz.
3. Select the "Channel Power ACLR" measurement function from the "Select Mea-
surement" dialog box.
4. Set the "W-CDMA 3GPP REV" standard for adjacent channel power measurement
in the "ACLR Setup" dialog box.
Channel Power and Adjacent-Channel Power (ACLR) Measurement