Rapidlab 1200 Operator’s Guide: Troubleshooting 6-19
02087462 Rev. V
c. Insert the clot clearing tool in the place of the pO
2
sensor.
Figure 6-5 Clot Clearing Tool inserted in pO
2
Sensor Slot
d. Press the black lever (item 2, Figure 6-5), located below the black
spring-loaded latch assembly, to reapply pressure to the sensor assembly.
3. Fill the syringe with water.
4. Insert the syringe in clot clearing tool (item 1, Figure 6-6).
Figure 6-6 Syringe inserted in Clot Clearing Tool
a. If using a Luer Lock syringe, thread the syringe into the clot clearing tool
to ensure the syringe is secured firmly to the clot clearing tool.
1 clot clearing tool
2 black lever
1 Clot clearing tool
2 Syringe