EasyManua.ls Logo

Agilent Technologies 7500 ICP-MS

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-3
Agilent 7500 ICP-MS ChemStation Operators Manual
Configuration
ICP-MS Configuration Dialog Box
2 Set the Offline Instrument check box to OFF (do not check this box).
On-line is the state in which data is transferred via the LAN or GPIB interface. In
this mode, you use ChemStation to control the operation , tuning and data analysis
Off-line, on the other hand, is the state in which ChemStation is used alone without
transferring data between ChemStation and the instrument, for example, to analyze
existing data.
To use ChemStation off-line, set the Off-line Instrument check box to ON (Check
this box).
3 Set the ICP-MS address in the GPIB area.
The initial value is set to 20. If the Offline Instrument box is checked, this value is
not used.
4 Set the Remote Start to start the data acquisition by an external signal.
Setting this mode allows the start of data acquisition using an external signal, e.g.
Laser Ablation System.
Don’t Use

Table of Contents

Related product manuals