EasyManua.ls Logo

Agilent Technologies 7500 ICP-MS

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-4
Agilent 7500 ICP-MS ChemStation Operators Manual
Configuration
Select this if you do not want to use the remote start mode.
Wait until Ready Signal
Select this if you want to use the remote start mode.
5 Set the sample introduction.
CAUTION RS-232 Configuration is necessary when ISIS or Autosampler are used. For the
setting method, see “I/O Library Setup” in Chapters 18 or 19.
Type
None : Select if you are not using the peristaltic pump, ISIS,
or LC/GC/Laser Ablation.
ISIS : Select if you are using the ISIS for data aqusition
Peristaltic Pump : Select if you are using only the instrument pump for
sample uptake and/or drain
LC/GC/laser abrasion: Uses LC/GC/laser abrasion.
Link to Agilent LC/GC ChemStation checkbox/Instrument# combo box:
Selecting LC/GC/Laser Ablation as Type enables the Link to Agilent LC/GC
ChemStation checkbox. Select this checkbox to use the LC/GC ChemStation.
Selecting the checkbox also enables the Instrument# combo box. Select the
number of the instrument you wish to control.

Table of Contents

Related product manuals