EasyManuals Logo

Agilent Technologies 7500 ICP-MS User Manual

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
3-5
Agilent 7500 ICP-MS ChemStation Operators Manual
Startup, Shutdown and Status
6 Start up the printer and computer.
Refer to the printer and computer user manuals for more information.
7 Start the ChemStation software.
Refer to Chapter 1, “Using the Agilent 7500 ICP-MS ChemStation Software” for
information on exiting the software. The ICP-MS Top window appears.
The ICP-MS Top Window
CAUTION When the ChemStation software starts running immediately after turning on the main
switch of the ICP-MS instrument, the mainboad of the instrument sometimes hangs
up and becomes unavailable to control the instruments. ChemStation software should
be started more than 40 seconds after turning on the main switch of the instrument.
8 Select Instrument>>Instrument Control.
The Instrument Control window appears. The title bar shows that the instrument is
in Shutdown mode. The LED located on the top right corner of the right top cover is
off during Shutdown mode. Continue with “Starting from Shutdown Mode”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals