EasyManuals Logo

Agilent Technologies 7500 ICP-MS User Manual

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #562 background imageLoading...
Page #562 background image
18-32
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Installing the Agilent 7500 ChemStation Software (Windows XP)
14 The message asking for the Agilent 7500 to be placed in Shutdown mode is
displayed. Click OK.
NOTE If the instrument is not in shutdown mode, press the shutdown switch on the
underside of the right instrument cover. The instrument transfers to shutdown mode
(it takes approximately 5 minutes). When it is in shutdown mode, the status LED
indicator on the cover turns off.
15 Reboot the computer.
16 Install the Patch File if necessary. Refer to page 18-35.
17 Select Start>>All Programs>>ICP-MS ChemStation>>Edit hosts. Enter the IP
Address and the Host Name of Agilent 7500 which is setup in the Bootp Service.
Save the hosts file.
18 Reboot the computer.
19 Turn off the power of the Agilent 7500. Confirm the cable connection between
the Agilent 7500 and the computer. Then turn on the Agilent 7500.
CAUTION After restarting the computer, turn off and turn on the Agilent 7500.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals