3-18
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Startup, Shutdown and Status
Meter Control Panel Values
Meter Typical Range
Shut Down M ode Standby M ode Analysis M ode
IF Press --- --- 250 to 490 (1 RP) *
180 to 300 (2 RPs)
Backing Press --- 0.3 to 5 250 to 490 (1 RP) *
180 to 300 (2 RPs)
Analyzer Press --- 3x10
-5
to 6x10
-4
3x10
-4
to 2x10
-3
TPI Revolution 0 65 to 100 95 to 100
TPA Revolution 0 65 to 100 95 to 100
Water WC/IF 0.00 0.00 1.1 to 2.0
Water RF/TP 0.00 0.00 1.1 to 3.0
Water Temp 15 to 25 15 to 25 15 to 25
Inlet Temp Room Temperature Room Temperature 25 to 35
Vent Temp Room Temperature Room Temperature 45 to 55
S/C Temp (L) Room Temperature Room Temperature 2
Carrier Gas Press 0 0 450 to 600 (BB) **
300 to 400 (CF)
200 to 300 (CN)
OP Gas Press 0 0 Depends on
app
li
cat
i
on
OP Gas Tank 0 0 Depends on
P
ress app
li
cat
i
on
Argon Gas Tank 700 700 700
Press

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals