EasyManuals Logo

Agilent Technologies 7500 ICP-MS User Manual

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
3-21
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Startup, Shutdown and Status
The Nebulizer Test dialog box will be displayed.
Nebulizer Test dialog box
3 Select the type of nebulizer from the Nebulizer Type area.
4 Adjust the source pressure of the argon gas to fit within the appropriate range
using a regulator.
The appropriate range will vary with the type of nebulizer and main unit.
(x denotes a numeral)
5Click Run.
The source pressure of the argon gas will be checked.
Nebulizer type Main unit type
G327xA G315xA/B
Flow rate for pressure
adjustment
Babington 675 - 725 kPa
(700 kPa + - 3.5%)
675 - 725 kPa (700 kPa +- 3.5%) 1.0 ml/min
Concentric 482 - 518 kPa (500 kPa +- 3.5%) 1.0 ml/min
Micro Flow(100) 337 - 363 kPa (350 kpa +/- 3.5%) 0.8 ml/min
Micro Flow(20) 337 - 363 kPa (350 kpa +/- 3.5%) 0.8 ml/min
MicroMist 675 - 725 kPa (700 kPa +- 3.5%) 1.0 ml/min

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals