Agilent 7500 ICP-MS ChemStation Operators Manual
Startup, Shutdown and Status
If the average value of the source pressures of the argon gas is 100 kPa or less, a
message indicating that the test has failed will be displayed in the Current Situation
area. Check to see whether the nebulizer is leaking.
This dialog box appears when there is a leak in the nebulizer.
If the average value of the measured flow rates is “set flow rate value - 0.1” or less
or the average value of the carrier gas pressures exceeds “argon tank pressure by
0.95,” a message indicating that the test has failed will be displayed in the Current
Situation area. Check to see whether the nebulizer is clogged.
This dialog box appears when there is a clog in the nebulizer.
7Click Close.
The Nebulizer Test dialog box will be closed.