Commissioning
82/198 CCD CAM 100 – Installation and Operating Manual
8.6.3 Selecting the Calibration Method
The third page of the menu (Calibration Method) is used to select
the method for the calibration of the CCD camera.
Fig.49: Calibration Method menu
You can select from the following settings:
Setting Description Remark
Off The CCD camera will not be calibrated.
Measurement inaccuracies may occur
due to the optical characteristics of the
camera lens.
Factory
The device uses the values from the calibration of
the CCD camera made in the factory for
compensating the measurement inaccuracies.
Use this calibration method, if no
calibration template is available.
High Res.
The calibration of the CCD camera is carried out
using a calibration template.
Use this calibration method for
applications requiring high measuring
accuracy.
You can find a detailed description of the settings in chapter
Selecting the Calibration Method, page 156.