EasyManua.ls Logo

HP E4418B - Page 111

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
prog.book : chapter1.fm 87 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Status Reporting
HP E4418B/E4419B Programming Guide 1-87
Device Status Register
The device status register set contains bits which give device dependent
information.
The following bits in these registers are used by the power meter:
1. HP E4419B only
Bit
Number
Decimal
Weight
Definition
0 - Not used
1 2 Channel A sensor connected
2 4 Channel B sensor connected
1
3 8 Channel A sensor error
4 16 Channel B sensor error
1
5 32 Channel A sensor Front/Rear
6 64 Channel B sensor Front/Rear
1
14 16384 Front Panel key press