prog.book : ch9sens.fm 31 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:FDOFfset|GAIN4[:INPut][:MAGNitude]?
HP E4418B/E4419B Programming Guide 9-31
[SENSe[1]]|SENSe2:CORRection:FDOFfset|GAIN4[:INPut][:MAGNitude]?
This command is used to return the frequency dependent offset currently
being applied.
Syntax
Example
Reset Condition
On reset, the frequency dependent offset is not affected.
CORR:GAIN4? This command queries the current
frequency dependent offset being
applied to channel A.
SENS
1
:GAIN4
CORR
SENS2
:
:FDOFfset
:INP
:MAG
?