EasyManuals Logo

HP E4418B Programming Guide

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #352 background imageLoading...
Page #352 background image
prog.book : ch9sens.fm 51 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
HP E4418B/E4419B Programming Guide 9-51
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
This command is used with the HP E-series power sensors to select one of
two power ranges.
•If0 is selected, the power sensor’s lower range is selected.
•If1 is selected, the power sensor’s upper range is selected.
Setting a range with this command automatically switches
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO to OFF.
Syntax
Example
Reset Condition
On reset, the upper range is selected.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe?
The query enters a 1 or 0 into the output buffer indicating the status of
the power sensor’s range.
1 is returned when the upper range is selected.
0 is returned when the lower range is selected.
Query Example
Error Message
This command is used with the HP E-series power sensors. If one is not
connected the error -241, “Hardware missing” occurs.
POW:AC:RANG 0 This command sets the power
sensor to it’s lower range.
POW:AC:RANG? This command queries the current
setting of the power sensor range.
SENS
1
:AC
POW
SENS2
:
:RANG
Space
?
numeric_value

Table of Contents