prog.book : ch15ieee.fm 15 Mon Feb 15 09:24:37 1999
IEEE488.2 Command Reference
*RCL <NRf>
HP E4418B/E4419B Programming Guide 15-15
*RCL <NRf>
The *RCL <NRf> (ReCaLl) command restores the state of the power meter
from the specified save/recall register. An instrument setup must have
been stored previously in the specified register.
Syntax
Parameters
Error Message
If the register does not contain a saved state error -224, “Illegal parameter
value” occurs.
Type Description/Default
Range of
Values
NRf The number of the register to be recalled. 1 to 10
*RCL
NRf
Space