prog.book : ch11syst.fm 19 Mon Feb 15 09:24:37 1999
SYSTem Subsystem
SYSTem:COMMunicate:SERial:TRANsmit:ECHO ON|OFF
HP E4418B/E4419B Programming Guide 11-19
SYSTem:COMMunicate:SERial:TRANsmit:ECHO ON|OFF
This command allows the power meter to be controlled by a dumb
terminal that may require its transmitted characters to be echoed. If ON is
specified, data received by the receiver is transmitted back to the sender
(echo on). If OFF is specified, data received by the receiver is not
transmitted back to the sender (echo off).
Syntax
Example
Reset Condition
On reset, the echo selection is unaffected.
Query
SYSTem:COMMunicate:SERial:TRANsmit:ECHO?
The query returns the current setting of the transmit/receive word length.
ON is returned if the power meter is set to return received data
back to the sender (echo on).
OFF is returned if the power meter is not set to return received
data back to the sender (echo off).
SYST:COMM:SER:TRAN:ECHO ON This command sets the power
meter to transmit any data
received back to the sender.
SYST
:COMM
:SER
:ECHO
:TRAN
Space
?
ON
OFF

Table of Contents