prog.book : ch4cal.fm 9 Mon Feb 15 09:24:37 1999
CALibration Subsystem
CALibration[1|2]:ECONtrol:STATe <Boolean>
HP E4418B/E4419B Programming Guide 4-9
CALibration[1|2]:ECONtrol:STATe <Boolean>
This command is used to enable and disable the rear panel TTL zero/cal
inputs. The TTL inputs provide an external means to initiate ZERO and
CAL cycles.
Syntax
Example
Reset Condition
On reset, the TTL zero/cal inputs are disabled.
Query
CALibration[1|2]:ECONtrol:STATe?
The query enters a 1 or 0 into the output buffer indicating whether the
TTL inputs are enabled or disabled.
• 1 is returned if the TTL inputs are enabled.
• 0 is returned if the TTL inputs are disabled.
Query Example
CAL1:ECON:STAT 1 This command enables the rear
panel TTL inputs. Note that
enabling the TTL inputs is not
channel dependent and CAL1 and
CAL2 have the same effect.
CAL1:ECON:STAT? This command queries whether or
not the TTL inputs are enabled
CAL
1
2
0|OFF
1|ON
Space
?
:ECON
:STAT