CALCulate Subsystem
CALCulate[1|2]:Limit:CLEar[:IMMediate]
HP E4418B/E4419B Programming Guide 3-11
CALCulate[1|2]:Limit:CLEar[:IMMediate]
This command is used to immediately clear the FCO (fail counter) of any
limit failures for the specified window. The FCO is used to determine the
results returned by the CALCulate[1|2]:LIMit:FAIL? query
Syntax
Example
CALC2:LIM:CLE:IMM This command clears the FCO for
the lower window.
CALC
1
2
:LIM
:CLE
:IMM