prog.book : ch10stat.fm 11 Mon Feb 15 09:24:37 1999
STATus Subsystem
STATus:OPERation
HP E4418B/E4419B Programming Guide 10-11
STATus:OPERation
The operation status register set contains conditions which are a part of
the operation of the power meter as a whole.
The following bits in these registers are used by the power meter:
These bits are set and cleared under the conditions shown on page 1-85.
Syntax
Bit
Number
Decimal
Weight
Definition
0 1 CALibrating Summary
1 - 3 - Not used
4 16 MEASuring Summary
5 32 Waiting for TRIGger Summary
6 - 9 - Not used
10 1024 SENSe Summary
11 2048 Lower Limit Fail Summary
12 4096 Upper Limit Fail Summary
13 to 15 - Not used (bit 15 always 0)