EasyManuals Logo

HP E4418B Programming Guide

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #320 background imageLoading...
Page #320 background image
prog.book : ch9sens.fm 19 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 Node
HP E4418B/E4419B Programming Guide 9-19
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 Node
This command node controls the pulse power measurement feature of the
power meter.
The commands associated with this node are:
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut]
[:MAGNitude] <numeric_value>
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <Boolean>
Note You can use either DCYCLe or GAIN3 in these commands, both do
the same. Using GAIN3 complies with the SCPI standard whereas
DCYCle does not, but may make your program more
understandable.