EasyManuals Logo

HP E4418B Programming Guide

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #308 background imageLoading...
Page #308 background image
prog.book : ch9sens.fm 7 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <Boolean>|ONCE
HP E4418B/E4419B Programming Guide 9-7
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <Boolean>|ONCE
This command is used to enable and disable automatic averaging. ONCE
has no affect on the power meter.
When the auto filter mode is enabled, the power meter automatically sets
the number of readings averaged together to satisfy the averaging
requirements for most power measurements. The number of readings
averaged together depends on the resolution and the power level in which
the power meter is currently operating. Figure 9-1 lists the number of
readings averaged for each range and resolution when the power meter is
in auto filter mode.
Setting this command to ON automatically sets the
[SENSe[1]]|SENSe2:AVERage:STATe command to ON.
Figure 9-1: Averaged Readings
If [SENSe[1]]|SENSe2:AVERage:COUNt:AUTO is set to OFF, the filter
length is set by the [SENSe[1]]|SENSe2:AVERage:COUNt command.
Using the [SENSe[1]]|SENSe2:AVERage:COUNt command
automatically disables automatic averaging.
Auto averaging is enabled by the MEASure:POWer:AC? and
CONFigure:POWer:AC? commands.
10 dB
10 dB
Minimum Sensor Power
Maximum Sensor Power
Power Sensor
Dynamic Range
10 dB
10 dB
1234
1 8 128 128
Resolution Setting
Number of Averages
1 1 16 256
11232
11116
1118

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals