prog.book : ch3calc.fm 9 Mon Feb 15 09:24:37 1999
CALCulate Subsystem
CALCulate[1|2]:LIMit:CLEar:AUTO <Boolean>|ONCE
HP E4418B/E4419B Programming Guide 3-9
CALCulate[1|2]:LIMit:CLEar:AUTO <Boolean>|ONCE
This command is used to control when the FCO (fail counter) is cleared of
any limit failures. The FCO is used to determine the results returned by
the CALCulate[1|2]:LIMit:FAIL? query.
If ON is specified, the FCO is set to 0 each time a measurement:
is initiated using INITiate[:IMMediate].
is initiated using INITiate:CONTinuous ON.
is measured using MEASure?
is read using READ?
If OFF is specified, the FCO is not cleared by the above commands.
If ONCE is specified, the FCO is cleared only after the first initialization
then starts accumulating any limit failures.
Syntax
Example
Reset Condition
On reset, both windows are set to ON.
CALC1:LIM:CLE:AUTO 1 This command switches on
automatic clearing of the FCO for
the upper window..
CALC
1
2
0|OFF
1|ON
Space
?
:LIM
:CLE
:AUTO
ONCE