Power Meter Remote Operation
Using Sensor Calibration Tables
1-38 HP E4418B/E4419B Programming Guide
Selecting a Sensor Calibration Table
After you have created the sensor calibration table, you can select it using
the following command:
[SENSe[1]]|SENSe2:CORRection:CSET1[:SELect] <string>
When the table is selected, the power meter verifies the number of
calibration factor points defined in the sensor calibration table is one
parameter greater than the number of frequency points. If this is not the
case an error occurs.
To find out which sensor calibration table is currently selected, use the
query:
[SENSe[1]]|SENSe2:CORRection:CSET1[:SELect]?
Enabling the Sensor Calibration Table System
To enable the sensor calibration table, use the following command:
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe ON
If you set [SENSe[1]]|SENSe2:CORRection:CSET1:STATe to ON and
no sensor calibration table is selected error -221, “Settings conflict” occurs.