prog.book : ch7mem.fm 26 Mon Feb 15 09:24:37 1999
MEMory Subsystem
MEMory:TABLe:GAIN[:MAGNitude] <numeric_value>{,<numeric_value>}
7-26 HP E4418B/E4419B Programming Guide
MEMory:TABLe:GAIN[:MAGNitude] <numeric_value>{,<numeric_value>}
This command is used to enter calibration factors into the sensor
calibration table, or offsets into the frequency dependent offset table,
currently selected using MEMory:TABLe:SELect. Any previous
calibration factor list, or offset list is cleared before the new calibration
factors/offsets are stored.
A maximum of 81 parameters for sensor calibration tables and 80
parameters for frequency dependent offset tables can be sent with this
command. For sensor calibration tables only, the first parameter is the
reference calibration factor, each subsequent parameter is a calibration
factor point in the sensor calibration table.
Entries in the frequency lists correspond as shown with entries in the
calibration/offset factor lists.
For sensor calibration tables the number of frequency points must be one
less than the number of calibration factor data points. This is verified
when the sensor calibration table is selected using
SENSe:CORRection:CSET1:SELect <string>.
Syntax
Calibration Factor/Offset Frequency
Reference Calibration Factor
(Sensor Cal Table Only)
-
Calibration Factor/Offset 1 Frequency 1
""
Calibration Factor/Offset 80 Frequency 80
MEM
:TABL
:GAIN
:MAGN
numeric_value
Space
,
?

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals