EasyManua.ls Logo

HP E4418B - [Sense[1]]|Sense2:Correction:loss2:State <Boolean

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
prog.book : ch9sens.fm 35 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe <Boolean>
HP E4418B/E4419B Programming Guide 9-35
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe <Boolean>
This command is used to enable and disable a loss correction for the power
meter setup, for example cable loss. Since LOSS2 and GAIN2 form a
coupled system, the command is identical to
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe.
The [SENSe[1]]|SENSe2:CORRection:LOSS2 command should be used
to enter the loss value.
Syntax
Example
Reset Condition
On reset, loss correction is disabled.
Query
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe?
The query enters 1 or 0 into the output buffer indicating the status of the
loss correction.
1 is returned if loss correction is enabled.
0 is returned if loss correction is disabled.
Query Example
SENS2:CORR:LOSS2:STAT 1 This command enables loss
correction on channel B.
CORR:LOSS2:STAT? This command queries whether
the loss correction on channel A is
on or off.
SENS
1
:LOSS2
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?

Table of Contents

Other manuals for HP E4418B

Related product manuals