prog.book : ch9sens.fm 35 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe <Boolean>
HP E4418B/E4419B Programming Guide 9-35
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe <Boolean>
This command is used to enable and disable a loss correction for the power
meter setup, for example cable loss. Since LOSS2 and GAIN2 form a
coupled system, the command is identical to
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe.
The [SENSe[1]]|SENSe2:CORRection:LOSS2 command should be used
to enter the loss value.
Syntax
Example
Reset Condition
On reset, loss correction is disabled.
Query
[SENSe[1]]|SENSe2:CORRection:LOSS2:STATe?
The query enters 1 or 0 into the output buffer indicating the status of the
loss correction.
• 1 is returned if loss correction is enabled.
• 0 is returned if loss correction is disabled.
Query Example
SENS2:CORR:LOSS2:STAT 1 This command enables loss
correction on channel B.
CORR:LOSS2:STAT? This command queries whether
the loss correction on channel A is
on or off.
SENS
1
:LOSS2
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?