SYSTem Subsystem
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric_value>
HP E4418B/E4419B Programming Guide 11-3
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric_value>
This command sets the GPIB address of the power meter.
Syntax
Parameters
Example
Query
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess? MIN|MAX
The query returns the current setting of the GPIB address or the values
associated with MIN and MAX.
Query Example
Item Description/Default Range of Values
numeric_value A numeric value for the address. The
default value, DEF, is 13. The value for MIN
is 0. The value for MAX is 30.
0 to 30
DEF
MIN
MAX
SYST:COMM:GPIB:ADDR 13 This command sets the GPIB
address to 13.
SYST:COMM:GPIB:ADDR? This command queries the setting
of the GPIB address.
SYST
:COMM
:GPIB
:SELF
:ADDR
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX