EasyManuals Logo

HP E4418B Programming Guide

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #386 background imageLoading...
Page #386 background image
prog.book : ch11syst.fm 3 Mon Feb 15 09:24:37 1999
SYSTem Subsystem
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric_value>
HP E4418B/E4419B Programming Guide 11-3
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric_value>
This command sets the GPIB address of the power meter.
Syntax
Parameters
Example
Query
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess? MIN|MAX
The query returns the current setting of the GPIB address or the values
associated with MIN and MAX.
Query Example
Item Description/Default Range of Values
numeric_value A numeric value for the address. The
default value, DEF, is 13. The value for MIN
is 0. The value for MAX is 30.
0 to 30
DEF
MIN
MAX
SYST:COMM:GPIB:ADDR 13 This command sets the GPIB
address to 13.
SYST:COMM:GPIB:ADDR? This command queries the setting
of the GPIB address.
SYST
:COMM
:GPIB
:SELF
:ADDR
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals