prog.book : ch9sens.fm 12 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage[:STATe] <Boolean>
9-12 HP E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:AVERage[:STATe] <Boolean>
This command is used to enable and disable averaging.
Syntax
Example
Reset Condition
On reset, averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage[:STATe]?
The query enters a 1 or 0 into the output buffer indicating the status of
averaging.
1 is returned when averaging is enabled.
0 is returned when averaging is disabled.
Query Example
Error Message
If [SENSe[1]]|SENSe2:AVERage:STATe is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, “Settings
Conflict” occurs.
AVER 1 This command enables averaging
on channel A.
SENS2:AVER? This command queries whether
averaging is on or off for channel B.
SENS
1
:STAT
AVER
0|OFF
1|ON
Space
?