prog.book : chapter1.fm 64 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Triggering the Power Meter
1-64 HP E4418B/E4419B Programming Guide
Initiate State
If the trigger system is on the downward path, it travels directly through
the initiate state without any restrictions. If the trigger system is on the
upward path, and INIT iate:CONTinuous is ON, it exits downwards to the
event detection state. If the trigger system is on the upward path and
INITiate:CONTinuous is OFF, it exits upwards to the idle state.
Event Detection State
The trigger source specifies which event causes the trigger system to
travel through the event detection state. The trigger source is set with the
following command:
TRIGger:SOURce
There are three possible trigger sources.
BUS
The trigger source is the HP-IB group execute trigger (<GET>), a
*TRG command, or the TRIGger:IMMediate command.
HOLD
Triggering is suspended. The only way to trigger the power meter
is to send TRIGger:IMMediate.
IMMediate
The power meter does not wait for any event and immediately
travels through the event detection state.
Querying the Trigger Source
The trigger source is queried with the following command:
TRIGger:SOURce

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals