prog.book : ch4cal.fm 12 Mon Feb 15 09:24:37 1999
CALibration Subsystem
CALibration[1|2]:RCFactor <numeric_value>
4-12 HP E4418B/E4419B Programming Guide
CALibration[1|2]:RCFactor <numeric_value>
This command is used with HP 8480 series power sensors to set the
reference calibration factor of the specified channel. Reference calibration
factors can also be set using sensor calibration tables. The power meter
uses the most recently set reference calibration factor.
Syntax
Parameters
Example
Reset Condition
On reset, the reference calibration factor is set to 100%.
Item Description/Default Range of Values
numeric_value A numeric value. The default value,
DEF, is 100%. The value for MIN is 1%.
The value for MAX is 150%.
1.0 to 150.0PCT
DEF
MIN
MAX
CAL1:RCF 98 This command enters a reference
calibration factor of 98% to
channel A.
CAL
1
2
:RCF
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space