prog.book : ch9sens.fm 37 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed] <numeric_value>
HP E4418B/E4419B Programming Guide 9-37
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed] <numeric_value>
This command is used to enter a frequency. If the frequency does not
correspond directly to a frequency in the sensor calibration table, the
power meter calculates the calibration factor using linear interpolation.
For HP 8480 series power sensors the power meter uses linear
interpolation to calculate the calibration factor for the frequency entered if
[SENSe[1]]|SENSe2:CORRection:CSET:STATe is ON. For HP E-series
power sensors, the calibration factor is calculated using the sensor’s
calibration factor data held in its EEPROM.
Syntax
Parameters
1. The following measurement units can be used:
•Hz
• kHz (10
3
)
• MHz (10
6
)
• GHz (10
9
)
Item Description/Default Range of Values
numeric_value A numeric value for the frequency. The
default value, DEF, is 50 MHz. The value
for MIN is 1 kHz. The value for MAX is
999.999 GHz. The default units are HZ.
1 kHz to
999.999 GHz
1
DEF
MIN
MAX
:CW
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
SENS
1
FREQ
SENS2
:
:FIX

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals