prog.book : ch11syst.fm 16 Mon Feb 15 09:24:37 1999
SYSTem Subsystem
SYSTem:COMMunicate:SERial:TRANsmit:BAUD <numeric_value>
11-16 HP E4418B/E4419B Programming Guide
SYSTem:COMMunicate:SERial:TRANsmit:BAUD <numeric_value>
This command sets the baud rate for both the transmitter and the
receiver. The baud rate for the transmitter and receiver are tied together
and can either be set by this command or the equivalent receiver
command [RECeive]:BAUD.
Syntax
Parameters
Example
Reset Condition
On reset, the baud rate is unaffected.
Item Description/Default Range of Values
numeric_value A numeric value for the baud rate. The
default value, DEF, is 9600. The value for
MIN is 50. The value for MAX is 115200.
50, 75, 110, 150,
300, 1200, 1800,
2000, 2400, 3600,
4800, 7200, 9600,
19200, 38400,
57600, 115200
DEF
MIN
MAX
SYST:COMM:SER:TRAN:BAUD 38400 This command sets the transmitter
and receiver baud rate to 38400.
SYST
:COMM
:SER
:TRAN
:BAUD
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals