Power Meter Remote Operation
Using Frequency Dependent Offset Tables
1-42 HP E4418B/E4419B Programming Guide
Editing Frequency Dependent Offset Tables
It is not possible to create any additional frequency dependent offset
tables. However, the 10 existing ones can be edited using the MEMory
subsystem. To do this:
1. Select one of the existing tables using:
MEMory:TABle:SELect <string>.
For information on naming frequency dependent offset tables see
“Naming Frequency Dependent Offset Tables”, on page 1-44. For
information on the current names which you can select refer to
“Listing the Frequency Dependent Offset Table Names”, on
page 1-43.
2. Enter the frequency data using:
MEMory:TABle:FREQuency <numeric_value>
{,<numeric_value>}
3. Enter the offset factors as shown in the table below using:
MEMory:TABle:GAIN <numeric_value>
{,<numeric_value>}.
4. If required, rename the frequency dependent offset table using:
MEMory:TABLe:MOVE <string>,<string>. The first <string>
parameter identifies the existing table name, and the second
identifies the new table name.
Frequency Offset
Frequency 1 Offset 1
Frequency 2 Offset 2
""
Frequency n Offset n

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals