CALCulate Subsystem
CALCulate[1|2]LIMit:FAIL?
3-12 HP E4418B/E4419B Programming Guide
CALCulate[1|2]LIMit:FAIL?
This query enters a 1 or 0 into the output buffer indicating whether there
have been any limit failures for the specified window. A limit failure is
defined as CALC[1|2]:LIMit:FCO? being non-zero. The FCO (fail
counter) can be zeroed using the CALC[1|2]:LIMit:CLEar command.
1 is returned when one or more limit failures have occurred.
0 is returned when no limit failures have occurred.
Syntax
Example
Reset Condition
On reset, the buffer is set to zero for both windows.
CALC1:LIM:FAIL? This command queries if there
have been any limit failures on the
upper window.
CALC
1
2
:LIM
:FAIL
?