EasyManuals Logo

HP E4418B Programming Guide

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
prog.book : ch2meas.fm 22 Mon Feb 15 09:24:37 1999
MEASurement Instructions
FETCh[1|2][:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
2-22 HP E4418B/E4419B Programming Guide
FETCh[1|2][:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window’s measurement function to single
channel with relative mode off, recalculates the measurement and places
the result on the bus. The result is a power based measurement and is
expressed in the units defined by UNIT[1|2]:POWer.
Syntax
Parameters
Refer to “The Optional Parameters”, on page 2-3 for additional details on
the parameters in this command.
Item Description/Default Range of Values
expected_value
(for the
expected
power level)
The expected power level
parameter can be set to DEF or a
numeric value. If a value is
entered it should correspond to
that set by CONFigure otherwise
an error occurs.The units of
measurement are dBm and W. The
default units are defined by
UNIT:POWer.
sensor dependent
DEF
1
resolution A numeric value for the resolution.
If it is unspecified the current
resolution setting is used. If a
value is entered it should
correspond to the current
resolution setting otherwise an
error occurs.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1
FETC
1
2
:POW
:AC
?
:SCAL
Space
expected_value
DEF
resolution
,
DEF
source list
,

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals