prog.book : ch2meas.fm 8 Mon Feb 15 09:24:37 1999
MEASurement Instructions
CONFigure[1|2] Commands
2-8 HP E4418B/E4419B Programming Guide
CONFigure[1|2] Commands
The CONFigure commands are used on the specified window to set,
• the expected power level value of the measurement.
• the resolution of the window.
• the channel(s) on which the measurement is to be made.
The CONFigure commands do not make the power measurement after
setting the configuration. Use READ?, or alternatively use INITiate
followed by a FETCh? to make the measurement.
The CONFigure command also does the following to the channel(s) which
are in the specified window (the channel(s) in the window are specified in
the <source list> parameter):
INITiate:CONTinuous OFF Sets the power meter to make one trigger
cycle when INITiate is sent.
TRIGger:SOURce IMMediate Sets the power meter to make the
measurement immediately a trigger is
received.
TRIGger:DELay:AUTO ON Enables automatic delay before making the
measurement.
SENSE:AVERage:COUNt:AUTO ON Enables automatic filter length selection.
SENSE:AVERage:STATe ON Enables averaging.