prog.book : ch15ieee.fm 22 Mon Feb 15 09:24:37 1999
IEEE488.2 Command Reference
*TRG
15-22 HP E4418B/E4419B Programming Guide
*TRG
The *TRG (TRiGger) command triggers all channels that are in the
“wait-for-trigger” state. It has the same effect as Group Execute Trigger
(GET).
Using the *DDT command may change the function of the *TRG command.
Syntax
Error Message
If TRIGger:SOURce is not set to BUS then error -211, “Trigger ignored”
occurs.
If the power meter is not in the “wait-for-trigger” state then error -211,
“Trigger ignored” occurs.
*TRG