prog.book : ch2meas.fm 11 Mon Feb 15 09:24:37 1999
MEASurement Instructions
CONFigure[1|2][:SCALar][:POWer:AC]:RELative [<expected_value>[,<resolution>[,<source list>]]]
HP E4418B/E4419B Programming Guide 2-11
CONFigure[1|2][:SCALar][:POWer:AC]:RELative
[<expected_value>[,<resolution>[,<source list>]]]
This command is used to set the measurement function, range and
resolution of the specified window. It sets the measurement function to
single channel with relative mode on. The relative value used is that set
by the CALCulate:RELative:MAGNitude:AUTO command.
Syntax
Parameters
Refer to “The Optional Parameters”, on page 2-3 for additional details on
the parameters in this command.
Item Description/Default Range of Values
expected_value
(for the
expected
power level)
A numeric value for the expected
power level. The units of
measurement are dBm and W. The
default units are defined by
UNIT:POWer.
sensor dependent
DEF
1
resolution A numeric value for the resolution.
If unspecified the current
resolution setting is used.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1
CONF
1
2
:POW
:AC
:SCAL
:REL
Space
expected_value
DEF
resolution
,
DEF
source list
,