SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <Boolean>
HP E4418B/E4419B Programming Guide 9-17
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <Boolean>
This command is used to enable and disable the use of the currently active
sensor calibration (CSET1) or frequency dependent offset (CSET2) table.
When a table has been selected and enabled, the calibration factors/offsets
stored in it can be used by specifying the required frequency using the
[SENSe[1]]|SENSe2:FREQuency command.
When the CSET1 command is set to ON, the reference calibration factor is
taken from the sensor calibration table and is used during calibration.
Syntax
Example
Reset Condition
On reset, the sensor calibration table and frequency dependent offset
table are not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a
table is enabled or disabled.
• 1 is returned when the table is enabled.
• 0 is returned when the table is disabled.
CORR:CSET1:STAT 1 This command enables the use of
the currently active sensor
calibration table for channel A.
SENS
1
:CSET
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?
1
:CSET2

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals