prog.book : ch9sens.fm 18 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <Boolean>
9-18 HP E4418B/E4419B Programming Guide
Query Example
Error Message
If you attempt to set this command to ON and no table has been selected
using the
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]
command then error -221, “Settings conflict” occurs and
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe remains
OFF.
SENS2:CORR:CSET1:STAT? This command queries whether
there is currently an active sensor
calibration table for channel B.