prog.book : ch13unit.fm 6 Mon Feb 15 09:24:37 1999
UNIT Subsystem
UNIT[1|2]:POWer:RATio <ratio_unit>
13-6 HP E4418B/E4419B Programming Guide
UNIT[1|2]:POWer:RATio <ratio_unit>
This command is used on the HP E4419B to set the default units used
when inputting and outputting power ratio data.
UNIT1:POWer:RATio sets the ratio measurement units for the
upper window. In addition, it also sets the ratio suffix for any
SENSe1 command that requires but omits the ratio suffix.
UNIT2:POWer:RATio sets the ratio measurement units for the
lower window. In addition, it also sets the ratio suffix for any
SENSe2 command that requires but omits the ratio suffix.
Syntax
Parameters
Example
Reset Condition
On reset, the units are set to dB.
Item Description/Default Range of Values
ratio_unit The ratio measurement unit. The
default unit is dB.
DB
PCT
UNIT1:POW:RAT DB This command sets the ratio
measurement units for the upper
window and the default units for
any relevant SENSe1 command to
dB.
UNIT :RAT
1
2
:POW
ratio_unit
Space
?

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals