prog.book : ch9sens.fm 46 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:LIMit:LOWer[:DATA] <numeric_value>
9-46 HP E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:LIMit:LOWer[:DATA] <numeric_value>
This command is used to enter a value for the lower test limit for the
specified channel used in the [SENSe[1]]|SENSe2:LIMit:FAIL? test.
The value is entered in dBm. When the measured power is less than the
value specified in [SENSe[1]]|SENSe2:LIMit:LOWer:DATA,
[SENSe[1]]|SENSe2:LIMit:FAIL? reports a fail. When the measured
power is greater than or equal to the limit, a fail is not reported.
Syntax
Parameters
Example
Item Description/Default Range of Values
numeric_value A numeric value. The default value,
DEF, is -90.00 dBm. The value for MIN
is -150 dBm. The value for MAX is
+230 dBm.
-150 to +230 dBm
DEF
MIN
MAX
SENS2:LIM:LOW:DATA MIN This command enters a lower limit
for channel B of -150 dBm.
:LOW
Space
?
:DATA
numeric_value
DEF
MIN
MAX
MIN
MAX
LIM
Space
SENS
1
SENS2
:

Table of Contents