prog.book : chapter1.fm 6 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Configuring the Remote Interface
1-6 HP E4418B/E4419B Programming Guide
Table 1-1: HP 437B Command Summary
Command Description
CL
Calibrate
*CLS
Clear all status registers
CS
Clear the status byte
CT
Clear sensor table
DA
1
All display segments on
DC0
Duty Cycle off
DC1
Duty Cycle on
DD Display disable
DE Display enable
DF
Display enable
DN
1
Down arrow key
DU
1
Display user message
DY
Enter duty cycle
EN
Enter
ERR?
Device error query
*ESR?
Event status register query
*ESE
Set event status register mask
*ESE?
Event status register mask query
ET
Edit sensor table
EX
Exit
FA
Automatic filter selection
FH
Filter hold
FM
Manual filter selection
FR
Enter measurement frequency
GT0
Ignore GET bus command
GT1
Trigger immediate response to GET
GT2
Trigger with delay response to GET
GZ
Gigahertz
HZ
Hertz
ID
Identification query
IDN?
Identification query
KB
Enter measurement cal factor