STATus Subsystem
STATus:OPERation:MEASuring[:SUMMary]
10-14 HP E4418B/E4419B Programming Guide
STATus:OPERation:MEASuring[:SUMMary]
The operation status measuring summary register set contains
information on the measuring status of the power meter.
The following bits in these registers are used by the power meter:
These bits are set when the power meter is taking a measurement.
These bits are cleared when the measurement is finished.
Syntax
Bit
Number
Decimal
Weight
Definition
0 - Not used
1 2 Channel A MEASuring Status
2 4 Channel B MEASuring Status (HP E4419B only)
3-15 - Not used
STAT
:OPER
:MEAS
:SUMM